Six Channel ECG Machine 10.0056 PRODUCT INFORMATION: |
 |
Large LCD Display Waveform & Manu
Six channel and Multi Form Printing
12 Lead Simultaneous Acquisition
Measurement & Interpretation
RS232 interface, software is Optional
16 set data and analysis result memory Is Optional
Features
Large LCD display six-channel ECG waveform, menus and parameter settings.
Standard six channel / multi printing forms available. Print date, time, parameters including paper speed, sensitivity, filter status.
12 leads simultaneously acquistion.
Measurement of HR, PR interval, QRS duration, QT/QTC interval, P/QRS/T axis, RV5/SVI and printing out measurement and result.
High resolution thermal array printer show clear and accurate waveform.
Controlling baseline position and drift is automatic.
Powerful digital filter can inhibit AC & EMG interference effectively.
AC and built in rechargeable battery power supply, 50 examinations available when the unit is fully charged.
Technical Specifications
Lead Standard 12 leads simultaneous acquisition
Operating Mode Automatic or manual
Recording System High resolution thermal printer
Input Circuit Isolated
Input Impedance > 50MÙ
Input Circuit Current < 50 nA
Patient Current Leakage < 10ìA
Sensitivity 5, 10, 20 mm/mV
Depolarisation Voltage + 300 mV
CMRR > 100 dB
Calibration Voltage 1 mV+3%
Noise Level <15 ìVp-p
Time Constant > 3.2 seconds
Frequency Response 0.05-150 Hz (-3 dB)
A/D Converter 12 bit
Sampling Rate 1 ms
Paper Speed 25, 50 mm/s
Recording Paper Rolling 112 mm x 27m, folding 112 mm x 75m
Display LCD
Safety Standard IEC Class I, type CF Protection from defibrillator & pacemaker
Power Supply AC: 220V+ 10%, 50~60 Hz; 110, 50 Hz DC: Built in rechargeable battery
Dimension 32.7 x 25.2 x 7.8 cm
Weight Net weight: 3.5 kg (including battery)
Gross weight: 7 kg
Dimensions when packed 37 x 31 x 25 cm
Standard Accessories
1 pc Patient Cable 6 pcs/set Chest Electrodes
4 pcs/set Lim Electrodes 1 pc Battery (Fixed in)
1 pc Power Cable 1 pc Grounding Cable
1 pc Paper Shaft 1 pc Roll Paper
3 pcs Fuse 1 Operation Manual